Medical Analyzer Housing: Cosmetics, Coating and Assembly

Project case study · October 6, 2026All Case Studies

A non-implantable instrument cover needed acceptable appearance around the display, a controlled coated surface and a consistent fit with the lower housing.

This anonymized summary uses owner-confirmed project records. Its measurements describe the reported component checks, not a general molding capability or finished-device approval.

Product design illustration of a tabletop analyzer upper housing and lower shell
Product design illustration, not a photograph of delivered parts. Geometry and color are illustrative.

Project Overview

Define the cover as a component of the instrument

The upper housing carried the instrument's external cosmetic surface and display mounting interface. FECISION's recorded responsibilities included DFM, review of the tooling proposal, molding-process development, appearance criteria, color control and assembly-fit evaluation. A partner factory manufactured the mold. This case does not describe a complete instrument development or regulatory approval project.

Project scope

ApplicationUpper cover and display support for a non-implantable tabletop analyzer
MaterialPC/ABS; final purchase designation reported as SABIC CYCOLOY C6600-701
ToolSingle-cavity tool with side gate and three slides
Released drawingRevision C, June 10, 2026
Production startJuly 15, 2026
Project rolesFECISION: DFM, tool coordination, process development and acceptance planning; partner factory: mold manufacture

Manufacturing Challenge

Three issues needed separate evidence

Product design illustration of the housing interior with a rib network and display mounting bosses
Structural illustration of the engineering review, not a before-and-after project photograph.

The T0 record identified a visible weld line around the display opening and sink on a broad cosmetic area. It also reported inconsistent performance of the internal conductive coating. The team needed to address appearance and coating preparation while preserving the display and lower-housing interfaces.

The initial notes associated the sink with a 3 mm wall region. They attributed the coating problem to surface contamination related to the trial material's release-agent history. That attribution guided the corrective work; the available summary is not a controlled study isolating every possible cause.

These issues could not be collapsed into a single acceptance number. Window flatness and assembled gap concerned geometry, visual inspection concerned the external surface, and coating measurements concerned a different functional layer. Each required a defined inspection method.

Engineering Response

Revise filling, wall distribution and surface preparation

The recorded tooling changes relocated the gate to move the weld line away from the designated cosmetic area. The team increased mold temperature for the subsequent trial and revised the broad wall region from 3.0 to 2.2 mm with a supporting rib network. These were project-specific changes, not a universal wall thickness or molding recipe.

For the coating route, the owner confirmed a change from the trial material designation CYCOLOY C6600-111 to the final purchasing designation CYCOLOY C6600-701, together with plasma treatment before coating. The resin designation is reproduced from the project confirmation; additive content and any material classification must follow the supplier's controlled documentation, not an inference from the suffix.

Because the material and preparation were changed together, this summary does not assign the result to either one alone. The revised configuration was assessed for appearance, dimensions, coating and assembly. The project record places trial-production acceptance on June 25, 2026, followed by the production start on July 15.

Recorded Results

Connect the reported values to their component checks

Product design illustration of an analyzer cover in an inspection fixture beside a lower shell and reference plaque
Inspection setup illustration, not a test photograph or measurement report.

The supplied inspection summary reported the following ranges and checks. The window flatness and assembly-gap ranges lie within their respective stated limits. The appearance record also reported no visible weld line or sink on the designated cosmetic surface under the agreed inspection.

The color record does not specify its color-difference formula or full instrument configuration. Its ΔE values are therefore presented as the project's recorded comparison, not a transferable acceptance method for another housing. Likewise, the coating resistance is a component-level result, not evidence of whole-device electromagnetic compatibility.

Owner-confirmed C02 inspection summary
CharacteristicRecorded requirementReported resultRecorded method
Display-window flatness≤ 0.20 mm0.10–0.18 mmCMM; nine points in the window region
Assembled housing gap0.30 ± 0.10 mm0.25–0.35 mmFeeler gauge; eight locations per part
Color differenceΔE ≤ 1.5ΔE 0.6–1.0D65 visual assessment and comparison with a reference panel
Coating sheet resistance≤ 0.5 Ω/□0.2–0.4 Ω/□Four-probe method; five internal coating locations per part

The intake lists a 100-piece trial-production batch and a cumulative 3,500 delivered housings through September 2026. Forecast monthly demand is not added to that delivered total. The supplied summary does not provide the population data needed to establish long-term process capability or to independently verify a production defect-rate claim.

Scope and Evidence

Preserve the limits of the completed review

The project record identifies extended disinfectant-wipe cycling and UV aging as unverified. The recorded application is indoor equipment; the case does not claim outdoor durability, sterilization suitability, biocompatibility, a flame classification or finished-device certification.

This anonymized summary uses project information supplied and confirmed by FECISION for public use. Original inspection reports and project photographs are not reproduced here. The supporting images are product design illustrations, not records of test results or customer-delivered parts.

The practical lesson is to keep cosmetic geometry, coating preparation and assembled fit in one coordinated review while retaining separate acceptance evidence. Passing one of those checks should not be used to close the others.

Related Engineering Guides

Prepare your own enclosure review

Bring the housing and its interfaces into the same quotation.

Send the controlled drawing, material, finish, counterpart revisions and approval requirements for your custom enclosure.